6.7 Calibrating the touch screen

1. Go to the main patient screen and stop ventilation.
P1: P1:
P2: P2:
Figure 6.10: "Access to screen calibration in patient mode" screen.
2. Press the
screen below appears.
<< press here
Figure 6.11: Calibration screens.
3. Press the first place indicated until you hear a beep.
4. Press the second place indicated until you hear a beep.
The touch screen is now calibrated.
The main patient screen reappears.
NOT013189-6
Monday 18
18 June
Monday
June
9:58:46
9:58:46
ACV
ACV
PS
PS
button and the Alarm Silence
CALIBRATION
Intern.
In
tern.
START
START
VENTIL
VENTIL
1s 1s
button simultaneously until the left-hand
CALIBRATION
press there >>
Patient use