ResMed Elisee 150 Clinical Manual page 119

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Message
MAINS MISSING.
REPLACE
BLOWER.
ADJUST
CLOCK.
Note: When the ventilator is powered by the mains power pack using an external power cord
(with the pack outside its compartment), this is treated as an external power supply. When this
power supply configuration becomes disconnected, it is the External Power Supply alarm which
is triggered rather than the Mains Power Supply alarm.
Warning
!
If the alarm persists after the remedies have been tried, stop using the ventilator and call
technical support.
NOT013189-6
Icon
Alarm priority
Low priority.
Low priority.
Low priority.
Cause
Remedy
No mains power
Check connection
in compartment
between mains
under device or at
power socket and
external power
mains power
socket.
pack, or if mains
power pack is
connected to
external power
socket, check this
connection, or
confirm
disconnection of
mains power by
pressing Alarm
Silence button.
Turbine
Return device to
maintenance.
ResMed service
centre.
Time setting lost.
Set the time.
Alarms
119/174