The test is very fast (about 1 minute) and consists of three stages:
Measuring the sensor offsets;
Measuring the inspiratory resistance of the circuit;
Measuring the expiratory resistance and calibrating the expiratory flow (for double circuit),
circuit compliances and O
Note: This test checks the O
at the bottom of the test results screen.
These measurements must be performed with all the accessories that will be used during
ventilation (expiratory valve, water trap, humidifier, etc.). Every time an accessory is added or
removed, this test must be repeated.
Caution
!
The event log records the time and results of each manual test.
Warning
!
During the test, the ventilator must not be connected to the patient.
If the test is unsuccessful, stop using the ventilator and contact the technician.
NOT013189-6
sensor gain.
2
sensor; if the sensor is found to be faulty, a message will appear
2
Clinical Configuration and Use