Implicit (calculated) parameters
Some parameters are adjustable, while others are calculated from the adjustable parameters. The
latter are implicit parameters (e.g. Ti and F are adjustable parameters; the implicit setting
Te = 60/F – Ti).
The adjustable parameters are: Max flow ( max), Ti, Te and I/E.
Note: Ti, I/E and max can be either adjustable or implicit parameters depending on the original
configuration selected for the volumetric modes (parameter options). If a beep sounds while the
user is setting the parameters, it means the I/E ratio has been inverted because of an
adjustment to one of the parameters.
Implicit parameter limits

Table 10.4: Implicit parameter limits

Max flow
( max)
V
Ti
I/E
Te
Total inspiratory pressure control
The threshold for total inspiratory pressure is: Pressure support + PEEP = 60cmH
PEEP = 60cmH
It is impossible to set a pressure higher than this value; even if a value higher than 60cmH
displayed, this limitation is always observed.
NOT013189-6
V
Minimum
Adult
10L/min
Paediatric
5L/min
5 L/min
1 / 0.4
0.45s
O, or Pinsp. + Precruit. + PEEP = 60cmH
2
V
Maximum
120L/min
40L/min
3.0s
O.
2
O, or Pinsp. +
2
O is
2
Appendices