Clean The Flow Generator; Record Clinical Settings - ResMed VPAP Auto 25 Service Manual

Hide thumbs Also See for VPAP Auto 25:
Table of Contents

Advertisement

1

Clean the flow generator

Before servicing the unit it should be wiped down using a lint free cloth and a cleaning agent with a
high alcohol content, e.g. isopropyl alcohol.
Warning
When using a cleaning agent with a high alcohol content refer to the appropriate Material
Safety Data Sheet to ensure safe handling of the cleaning agent.
!
Remove any stains from the cases. When removing any stains from the flow generator or mask
system use soap or detergent-based cleaning solutions. Remove excess and wipe dry after use.
Do not use thinners, trichlorethylene, bleach or other solvents. Do not apply oil to any component
of the flow generator.
2

Record clinical settings

1 If applicable, remove the front muffler assembly.
2 With the unit unplugged from the wall, remove the data card module.
3 Connect the ResScan USB adaptor to the flow generator. Press upwards and hold the button on
the adaptor, while pushing it onto the unit.
4 Turn on the flow generator and allow it to complete the self test.
5 Connect the flow generator to the PC via the USB cable.
3 General inspection and servicing