Table 5.1: Examples Of Log Events - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

The clinician's log sets out the following information, day by day:
Every triggering of an alarm;
The times when ventilation began and ended;
The type of power supply used;
The times when the ventilator was turned on and off.
23
cm
H2O
Monday
11 : 47: 14
90
10 : 06: 56
80
10 : 06: 54
70
10 : 06: 54
60
10 : 05: 55
50
9: 58: 43
40
HP
HP
9: 56: 37
30
9: 56: 04
20
9: 53: 07
10
9: 53: 06
<PE
<PE
0
9: 53: 06

Table 5.1: Examples of log events

Event displayed
11 : 47: 14
11 : 34: 14
11 : 02: 49
NOT013189-6
18 18
2007
2007
June
Start Ventil
Pb Silence
REPLACE O2 CELL
Pb Silence
REPLACE O2 CELL
Pb Silence
Pb Silence
MODE Change
Pb Silence
REPLACE O2 CELL
Pb Silence
MAXI P inspi
MINI
MINI
Vt
60 60
+ - al HP
0
Back
Back
1
1
1
1
0
1
1
Description of the event
The Maximum Pressure alarm was triggered at
11:47:14.
The Minimum Tidal Volume alarm stopped at 11:34:14.
The High Pressure alarm threshold was modified at
11:02:49. Its previous value was 60cmH
Press to access page for
previous day
Press to scroll up
Press to scroll down
O.
2
Clinical Configuration and Use
85/174

Advertisement

Table of Contents
loading