Testing settings alarms for use with a single limb circuit

Connect a single limb patient circuit fitted with a MAQUET test lung to the ventilator and perform
the manual test.
Set up ACV mode, non-invasive, adult ventilation.
Set the ventilation parameters as follows:
V
= 500mL;
T
PEEP = 0cmH
F = 15bpm;
Max flow ( ) = 23L/min (Ti = 1.3 or I/E = 1/2.0);
V
Trigger = no;
Flow shape = constant;
Start ventilation and perform the following operations.

Table 7.11: Testing settings alarms for use with a single limb circuit

Alarm
MAXI P expi.
MINI Vti.
MAXI Vti.
NOT013189-6
O;
2
Visual alarm
Action
status
Block the single
Flashing.
circuit valve
outlet.
Unblock the single
Steady.
circuit valve
outlet.
Press Alarm
Off.
Silence button.
Adjust alarm
Flashing.
upper threshold to
V
= 2000mL and
TE
lower threshold to
V
= 800mL.
TE
Adjust alarm
Steady.
lower threshold to
V
= 150mL.
TE
Press Alarm
Off.
Silence button.
Adjust alarm
Flashing.
upper threshold to
V
= 250mL.
TE
Adjust alarm
Steady.
upper threshold to
V
= 2000mL.
TE
Press Alarm
Off.
Silence button.
Audible alarm
Time taken for
status
alarm to trigger
Sounding.
1 cycle.
Off.
Off.
Sounding.
5 cycles.
Off.
Off.
Sounding.
5 cycles.
Off.
Off.
Alarms
125/174