If the ventilator is in the patient screens and delivering ventilation, the following screen appears.
cm
15
H
O
2
90
80
70
60
50
40
30
20
< PS
10
0
< PE
Figure 3.21: "Turning off the ventilator in patient mode during ventilation" screen (as Fig 3.19).
If the ventilator is in the clinical screens and delivering ventilation, the following screen appears.
cm
15
H
O
2
90
80
70
60
50
40
30
20
< PS
10
0
< PE
Figure 3.22: "Turning off the ventilator in clinical mode during ventilation" screen (as above).
In all three cases, press "Stop machine" to turn off the Elisée 150.
The flashing message "Stop in progress" is then displayed and the ventilator turns off. An audible
alarm will sound. Press the Alarm Silence button
Note: If the ventilator remains connected to mains power or an external power supply, it cannot
be turned off. It must first be disconnected from the power supply.
In the event of touch screen failure, it is possible to force the ventilator to stop in the following
manner:
1. Disconnect the ventilator from mains power or the external power supply;
2. Press the
3. Simultaneously press the Alarm Silence button
4. Hold both buttons down until the ventilator stops (10 seconds).
NOT013189-6
Pause (stop ventilation)
Stop machine
MENU
MENU
Configuration
Options
...
Pause (stop ventilation)
Stop machine
button on the right side of the ventilator;
Back
Back
to turn it off.
;