Connecting Accessories - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Connecting accessories

Note: Remember to perform a manual test after every change made to the patient circuit.
Antibacterial filter
Figure 3.12: Connecting an antibacterial filter.
The Elisée 150 can be fitted with an antibacterial filter or a heat and moisture exchange (HME)
filter. Either type of filter can be connected to the Y-coupler, to the insufflation cone or to the
exhalation cone.
Note: It is advisable to connect an antibacterial filter to the insufflation outlet or to connect an
HME filter between the Y-coupler and the patient.
The antibacterial filter needs to be fitted the right way round, as described in the manufacturer's
instructions.
Active humidification system
Figure 3.13: Connecting a humidifier.
The ventilator can also be combined with a humidification system to improve the humidity of the
air delivered to the patient.
Connect the heated base to a power supply in accordance with the manufacturer's instructions.
Caution
!
Assemble the humidifier in accordance with the manufacturer's instructions. Generally
speaking, the ventilator and expiratory valve should be placed above the humidifier and
water traps.
32/174
NOT013189-6