Replacing The O2 Cell - ResMed Elisee 150 Clinical Manual

Hide thumbs Also See for Elisee 150:
Table of Contents

Advertisement

Replacing the O
Caution
!
The replacement frequency for the O
given as an indication only and varies depending on the duration of ventilation, the oxygen
concentration used and the ambient temperature.
The O
cell is located at the rear of the ventilator.
2
To replace it:
1. Remove the protective cap from the O
Figure 8.4: Removing the protective cap from the O
2. Pull out the connector so that it does not obstruct the O
(see reference on page 149) to unscrew the O
with a new one.
Figure 8.5: Removing the O
Note: After opening the bag, wait 15 minutes before changing the O
Check the expiry date (month and year) on the label of the O
before the date on the label.
3. Screw the O
back on.
4. Perform a manual test to ensure the cell is functioning correctly (see "Manual test" on
page 65).
NOT013189-6
cell
2
cell.
2
cell back in place, plug the connector back in, and then put the protective cap
2
cell varies from 6 to 12 months. This frequency is
2
cell.
2
cell.
2
cell, then use the disassembly key
2
cell (see reference on page 149). Replace it
2
cell. The cell can be installed on or
2
cell.
2
Maintenance and Disinfection
131/174