Press "YES" to activate an option or "NO" to deactivate it.
To select the ACV parameter, press the Ti, max or I/E icon.
The icons for the selected options are highlighted.
Once all parameters are set, confirm by pressing "OK".
If OK is not pressed, none of the new settings will be saved.

Description of the parameter options

BTPS correction
When ventilating a patient, you must correct the BTPS (body temperature, ambient pressure,
saturated with water vapour). The correction is applied only to the measured expiratory volume
(V
). This option must be deactivated for ventilation not involving a patient. Measurements are
TE
taken in ATPD (ambient temperature and pressure, dry) conditions.
This option is activated by default. It is deactivated when the technical menu is accessed.
P2 activation
Allows you to program two ventilation options (P1 and P2). The patient can choose one of these
ventilation options. When the clinician configures a new patient ventilation with the P2 activation
option selected, the new configuration is saved in both P1 and P2.
Ti, max flow and I/E settings in volumetric (ACV) mode
This option allows the clinician to choose the set parameter to be used in volumetric ventilation.
The available options are inspiratory time (Ti), maximum flow ( max) and I/E ratio.
Note: Modifying any one of these parameters can change the ventilation settings.
92/174
PS
PS
Parameters option
BTPS correction
YES
YES
P2 activation
YES
YES
Select ACV parameter:
Ti Ti
Vmax
Vmax
Cancel
Cancel
info
NO
NO
O
O
NO
NO
K
K
I/E
I/E
V
V
NOT013189-6

Advertisement

Table of Contents
loading

Table of Contents