248476r4.book Seite 33 Freitag, 28. Oktober 2011 3:10 15

Setup menu

The Setup menu allows you to view and change therapy, alarm and device settings.

Accessing Clinical mode

CAUTION
For home treatment the device must be set to Patient mode
Only trained and authorised personnel must make changes to clinical settings.

Setup menu: Clinical Settings

On this screen, Clinical Settings are displayed according to the set therapy mode and can be
changed. Additional settings are located on the Advanced Settings screen.
.
To activate Clinical mode
, hold down
simultaneously for at least three seconds. You
will be prompted for how long the device should
stay in this mode. After the set period of inactivity
or next power cycle (power on/off), the device
automatically switches back to the patient mode
and the device beeps.
When Clinical mode is enabled, the lock symbol in
the header changes to unlocked and the device
beeps.