Turning the oxygen monitoring on and off
Turning the oxygen monitoring on
1. Turn off the ventilator (see "Turning off" on page 36).
2. Fit the oxygen sensor into its compartment (see "Replacing the O2 cell" on page 131).
3. Turn on the ventilator (see "Turning on" on page 35) and perform a manual test (see "Manual
test" on page 65).
The FiO
reading is now available (see "Measurements" on page 82) and the adjustable minimum
2
and maximum FiO
Turning the oxygen monitoring off
1. Turn off the ventilator (see "Turning off" on page 36).
2. Remove the oxygen sensor from its compartment and use the plug to seal the hole (see
"Replacing the O2 cell" on page 131).
3. The FiO
measurement is now no longer available (the value is replaced by dashes) and the
2
adjustable minimum and maximum FiO
displayed on the screen).
Note: When the sensor is removed, the R
Silence button
After removing the oxygen sensor, remember to insert the plug to prevent leaks.
NOT013189-6
alarm is activated.
2
2
once.
alarms are no longer activated (although still
alarm is triggered. Press the Alarm
EPLACE O 2 CELL
Installation
27/174