6. Fill in the Run Calibration menu as follows (see Figure 3-5):
Text enclosed in [brackets] denotes a custom user entry.
a. Test type: Alignment
b. Adjustment Tol % : [user entry]
c. Use Tol: Yes to use the Adjustment tol % from the last step or No
d. Head Relative to 4322: 0 if on the same surface. If not, measure the head-height
difference and enter it into the field. Use a positive (+) value if the UUT is above
the Controller and a negative (-) value if the UUT is below.
e. Reference SN: [serial number of the KN 3689]
f.
Addr: 10
g. Time/Date: [automatically populated]
h. Operator: [enter initials]
Figure 3-5. Alignment Menu Configuration
Figure 3-4. Module Selection
Note
Calibration and Alignment
Module Alignment Procedure
hbx067.jpg
hbx068.jpg
3
3-9