5.
Press [GUIDE].
Select [Dual Pointer], and press [START].
The dual pointer is will deactivate by pressing [STOP].
For LP-M5xx / LP-M2xx type:
Set "0mm" to the Z offset of the guide indication.
Place the power meter detector overlapping with the cross indicator of the
dual pointer.
Never place the detector where the point indicator is overlapping with the
point indicator, that indicates the focal point of the laser.
For LP-Sxxx type:
Place the power meter detector overlapping with the circle indicator of the
dual pointer.
Never place the detector where the point indicator is overlapping with the
circle indicator, that indicates the focal point of the laser.
6.
After setting the laser power meter, press [NEXT].
7.
Press [LASER CHECK].
Press [START] to radiate the laser and measure the
output power with the laser power meter.
Laser radiation is terminated by pressing [STOP].
WARNING
Reference
• For LP-M5xx-S / M2xx-S type, laser can not be emitted without opening the laser gate.
After starting the radiation, open the laser gate by external control signals.
Refer to the "Safety / Setup / Maintenance Guide" for detail.
• Be sure to wear protective goggles.
• During the radiation, the laser energy is concentrated to one point. Use due
caution with long period radiation, it may cause a fire or damage to the object.