 Following this double selection and confirmation via the Service
key, the service routine is activated.
11.10.2.3
Select parameters
If arrow keys are displayed in the selection fields once the required
Select parameter XG 3D / XG 3D ready / GALILEOS
service routine has been selected, you can use these arrow keys to
choose between different parameters.
You want to run service routine S017.6 to activate the remote control.
Once you have selected service routine S017.6, the code "00" is pre-
selected for the "Remote control disabled" option.
➢ Touch the + or - arrow keys (A) to select the code 01 (B) for the
"Remote control enabled" option.
 Once the selected parameter has been changed (in this case the
code for the activation of the remote control), the Memory key (C)
lights up.