S017: Configuration Service; 1S017: Test Step 2 - Sirona Orthophos XG 3D/Ceph Installation Manual

Table of Contents

Advertisement

11 Initial startup
11.10 Service routines for startup
SR*
S017
S017.2
S017.3
S017.4
S017.5
S017.6
S017.8
S017.12
S017.13
S017.14
S017.15
S017.17
S017.18
S017.21
130
11.10.3

S017: Configuration service

SA**
Function
S017 XG 3D
Unit configuration
Yes
Configuring the hardware version
Yes
Enter the country group code
Yes
Select a language
Yes
Select a language set
Yes
Enable/disable the remote control
Yes
Selecting the kV/mA level series
Yes
Enabling/disabling the preview image
Yes
Enable/disable the welcome screen
Yes
Enable/disable certain lines of the welcome screen
Yes
Activate/deactivate the acoustic signal for end of exposure
Yes
Trigger save/restore function for DX61 data
Yes
Activation of the occlusal bite block function (for occlusal bite block only)
NOTICE! This service routine may be selected only for units with an
occlusal bite block; otherwise, the unit will malfunction!
Yes
Configuration of angle pre-selection (temporomandibular joint) for the
TM1 and TM2 user programs
* SR=service routine, ** SA=security access
11.10.3.1
S017: Test step 2
Configuring the hardware version
S017_2 XG 3D / 3D ready
Selection field
1
* Factory setting
Installation Manual ORTHOPHOS XG 3D / Ceph
Code
Function
0001
Panoramic digital
0003
Panoramic digital, ceph left digital
0005
Panoramic digital, ceph right digital
0201
XG 3D + panoramic digital*
0203
XG 3D +
panoramic digital, ceph left digital
0205
XG 3D +
panoramic digital, ceph right digital
Sirona Dental Systems GmbH
63 03 452 D3352
D3352.031.05.08.02 10.2015

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents