Measured Variables and Operating Menus for the Standard Signal General
9.2

Changing the Measured Variable

WARNING!
Incorrect metering due to incorrect measured variable
Possible consequence: Fatal or serious injuries
The measured variable/measuring range of the sensor is essential for the meas‐
ured variable/measuring range!
If the measured variable/measuring range is changed, the sensor must be recali‐
brated
If the measured variable/measuring range is changed, the set points and limit
values will be changed to the associated starting values!
If the measured variable/measuring range is changed, the settings in all of the
menus must be checked
general setting
informationen
Fig. 48: Changing "Standard signal general" measured variable
For possible measured variables see
Identcode:
D1CBxxxxxxx
Vxxxxxxxxxx
Srnr: xxxxxxxxxx
Ä Table on page 105 .
Calib. timer
wash timer
measure change
add. functions
PAA
H2O2 (PER)
mA general
pH over mA
effective range
0...105 %
A0232_GB
103