Selecting the calibration
9.4.1
process for amperometric
measured variables
There are two calibration processes avail‐
able for calibrating the controller:

Calibrating the slope

n
Calibrating the zero point
n
Selecting the calibration process
1.
Continuous display ➨
CAL Cl
Last calibration 31.03. 2013 13:11:11
Slope
100 %
Zero point
4.00 mA
Slope calibration
Calibration of zero point
Fig. 56: [Chlorine] calibration menu
ð The calibration menu is dis‐
played.
2.
Use the arrow keys to select the
chosen menu item. Press
ð You can now start your chosen
calibration process.
9.4.2
Calibrating the slope
CAUTION!
Correct sensor operation / Run-in
period
Damage to the product or its sur‐
roundings
Correct measuring and metering
is only possible if the sensor is
working perfectly
Please read the operating manual
for the sensor
Please also read the operating
manuals for the fittings and other
components used
It is imperative that the run in
periods of the sensors are
adhered to
The run in periods should be
A1039
allowed for when planning com‐
missioning
It may take an entire working day
to run in the sensor
Calibration
85