Firmware Updates; With Signature Flowmeter; With 2160 Laserflow Module; Cleaning The Sensor Housing - Teledyne TIENet 360 LaserFlow Ex Installation And Operation Manual

Velocity sensor
Hide thumbs Also See for TIENet 360 LaserFlow Ex:
Table of Contents

Advertisement

TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 5 Maintenance and Optional Kits

5.2 Firmware Updates

5.2.1 With Signature
flowmeter
5.2.2 With 2160 LaserFlow
Module
5.3 Cleaning the Sensor
Housing
5.4 Accessing the
LaserFlow Ex's Nose
Piece
Tools required
5-2
When firmware updates become available, they can be down-
loaded from www.isco.com or via email from the factory.
The TIENet device's firmware is updated via the USB port on the
front panel of the Signature flowmeter. Step-by-step instructions
for updating the firmware can be found in Section 2 of the Sig-
nature user manual.
The TIENet device's firmware is updated via the "Update Isco
Instrument Software" tool from Flowlink for the 2160 module.
Step-by-step instructions for updating the firmware can be found
in the Help text of the update application.
The LaserFlow Ex sensor's outer housing and ultrasonic trans-
ducer can be cleaned with warm water, mild soap, and a soft,
lint-free cloth. Do not spray water on the laser window; install
the protective window cap prior to cleaning.
Never use ordinary cloth or abrasives to clean the LaserFlow
Ex as they may cause scratches.
Be sure the protective window cap is installed during cleaning,
storage, or shipment (refer to Section 1.6.1).
Several procedures for maintenance and options require
accessing the interior of the sensor by separating the nose piece
from the body, and subsequent reassembly. Refer back to this
section when performing these procedures.
Disconnect power from the LaserFlow Ex sensor before
opening the sensor housing.
• #2 Phillips straight screwdriver less than 4"
(right-angled if available)
CAUTION
CAUTION
WARNING