TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 2 Preparation and Installation
2.5.3 Sensor Positioning and
Requirements
a
a
Center of the pipe
Top of pipe inside wall
*
Distance from sensor tip to laser
beam exit point (vertex of 45° angle)
2-16
The figure on the left shows constants that serve as reference
points when planning the positioning of the sensor.
Figure 2-16, shows a typical LaserFlow Ex manhole installation.
For proper function, the requirements listed here must be fol-
lowed exactly.
Requirements:
• Permanent Installation: Install the wall bracket perpen-
dicular to the x-axis.
• Always center the LaserFlow Ex (and middle laser
beam) relative to the pipe walls, and always use the
attached circular bubble level for vertical alignment.
• Mount the LaserFlow Ex sensor with the bottom
parallel with the water surface.
• Align the bottom of the LaserFlow Ex with the top of the
inside wall of the inlet pipe, and as close as possible to
the pipe mouth without obstructing the laser.
To accomplish this, refer to the figure at left, Figure 2-16
on the next page, and the following important figures:
· The point at which the laser exits the sensor mouth is
3.3" (8.4 cm) from the front tip of the sensor, or the
center of the opening.
· The distance from the bottom of the LaserFlow Ex
sensor to the center of the coupling tube is 10.375"
(26.35 cm).
· The laser beam exits the sensor mouth at a 45°
angle.
(If the sensor bottom is parallel with the water
surface, the laser beam will also strike the water
surface at a 45° angle.)