Laser Velocity Measurement Point Selection - Teledyne TIENet 360 LaserFlow Ex Installation And Operation Manual

Velocity sensor
Hide thumbs Also See for TIENet 360 LaserFlow Ex:
Table of Contents

Advertisement

4.3.2 Laser Velocity
Measurement Point
Selection
Figure 4-13 Selection of velocity measurement points
TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 4 Setup with 2160 LaserFlow Module
Note
When sensor positioning is completed, securely tighten all
mounting hardware.
Once the LaserFlow Ex sensor is installed over the flow stream,
some installations may only require the center laser position for
velocity measurement.
For installations requiring multiple measurement points,
determine which points will be used. To do so, activate the 3
points and wait for all of them to display their readings. This
process may take up to 15 minutes.
Each point will display a velocity reading and time stamp, or
"Invalid." Deselect "Invalid" and undesired points. The fewer
points selected, the less time required to complete one velocity
measurement.
Select the "Select All" check box to activate the 3 points,
then click the "Apply" button.
Deselect "Invalid" and undesired points,
then click the "Apply" button.
4-15