3.4.3 Ultrasonic Level
Adjustment
Difference from bottom
(silt level)
360 Velocity
Laser On
Laser Off
Measurements are displayed in [UOM].
360 Velocity
Laser On
Laser Off
Measurements are displayed in [UOM].
Once the LaserFlow Ex sensor is installed over the flow stream,
measure the present liquid level from the bottom of the channel
(if there is a difference between the bottom of the channel and
zero level, save this value for reference in flow calculation) and
enter this value for Level, under Adjust options. Then highlight
Adjust and press Enter to confirm.
From this screen, you can also update the display to show the
current level of the stream.
Level
Note
1.00
0.00
Level adjustment of a newly installed and activated sensor
should be performed only after the sensor is allowed to stabi-
lize under site conditions (allow approximately one hour per 10
°F difference between storage and operating environments).
TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 3 Setup with Signature® flowmeter
Update
1.01
7/17/12
3:30:27 PM
0.99
7/17/12
3:31:52PM
Invalid
Update
1.01
7/17/12
3:30:27 PM
0.99
7/17/12
3:31:52PM
3-11