Figure 5-3 Cleaning the laser window: Moisten Kimwipe tissue w/ Rain-X
5.7 Installing the Remote
TIENet™ 310
Ultrasonic Level
Sensor
5-6
on the surface.
If the window is smudged, streaked, or still dirty, get a
fresh pre-packaged tissue and repeat.
4. Next, fold a clean, dry Kimwipe tissue four times to rein-
force its thickness, and gently clean the window surface
with downward strokes only, keeping the tissue
between your finger and the glass at all times.
5. Examine the window closely under good lighting from all
directions to ensure there are no smudges, streaks, or film
on the surface.
If the window is smudged, streaked, or still dirty, get a
fresh pre-packaged tissue and repeat.
6. Fold a fresh Kimwipe tissue as previously described and
spray the closed end of it with Rain-X.
7. Apply the Rain-X to the entire window surface using down-
ward strokes only.
8. After allowing 30 minutes for the Rain-X treatment to dry,
wipe the window with another dry Kimwipe tissue, as
described in step 4.
9. Carefully examine the window as previously described,
and continue polishing the surface with fresh dry tissues
until no streaks or film are visible.
Reassemble the nose piece, following the steps provided in
Section 5.4.2 Replacing the Nose Piece.
Refer to the TIENet™ 310 Ultrasonic Level Sensor Installation
and Operation Guide for details.
Spray this end
with Rain-X.