Level; Level Measurement Tab In Flowlink - Teledyne TIENet 360 LaserFlow Ex Installation And Operation Manual

Velocity sensor
Hide thumbs Also See for TIENet 360 LaserFlow Ex:
Table of Contents

Advertisement

4.2.1 360 Level

Ultrasonic Level
Adjustment
Difference from bottom
(silt level)
Figure 4-5 360 Level Measurement tab in Flowlink
The Measurement field displays the most recent level reading.
Once the LaserFlow Ex sensor is installed over the flow stream,
measure the present liquid level from the bottom of the channel
(if there is a difference between the bottom of the channel and
zero level, save this value for reference in flow calculation). Enter
this value in the Adjust level field. After the "Apply" button is
clicked, the level you entered will appear in the Measurement
field.
Level
Note
1.00
0.00
Level adjustment of a newly installed and activated sensor
should be performed only after the sensor is allowed to stabi-
lize under site conditions (allow approximately one hour per 10
°F difference between storage and operating environments).
TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 4 Setup with 2160 LaserFlow Module
4-5