Laserflow Ex Placement Using Slope Percentage Option - Teledyne TIENet 360 LaserFlow Ex Installation And Operation Manual

Velocity sensor
Hide thumbs Also See for TIENet 360 LaserFlow Ex:
Table of Contents

Advertisement

2%
slope
Figure 4-12 LaserFlow Ex placement using Slope
Percentage option
H - Optical Clarity
TIENet™ 360 LaserFlow Ex™ Velocity Sensor
Section 4 Setup with 2160 LaserFlow Module
Optical Clarity – Sense voltage detects condensation on the
window of the LaserFlow Ex sensor. When the sense voltage is
greater than
100 mV, the defogger is activated. The defogger is enabled from 1
to 5 minutes, depending on the severity of the fog.
Note
In locations with severe fogging potential, Teledyne ISCO rec-
ommends increasing battery capacity or monitoring the site
more frequently.
To enable Optical Clarity, click "Active". To disable Optical
Clarity, click "Passive". The default is "Active".
.
4-13