Operation
Using the 11970A, Q, U, V, and W Mixers with the X- Series Signal Analyzers (Option EXM)
Using the 11970A, Q, U, V, and W Mixers with the
X- Series Signal Analyzers (Option EXM)
Refer to
"Spectrum Analyzer Retrofit Requirements" on page 9
The following examples explain how to, connect the external mixers to the signal analyzer using a
diplexer, choose the band of interest, activate conversion loss correction data, and how to use the
signal-identification functions.
Set up the equipment
1. Connect the signal source, diplexer, and harmonic mixer to the signal analyzer as shown in
2-7.
The maximum torque for the SMA cables and adapters should not exceed 122
NOTE
N-cm (10 in-lb.)
Figure 2-7
External Mixer Setup
2. On the analyzer press
3. Select external mixing by pressing
Setup
.
4. The default is 11970A. To select Q, U, V or W bands, press
select the appropriate key corresponding to 11970 frequency band.
5. Press
twice to view the spectrum display.
Return
6. Tune the analyzer to the input signal frequency by pressing
34
Mode
,
Spectrum Analyzer
Input/Output
for configuration requirements.
.
,
,
External Mixer
External Mixer
Mixer Presets
FREQ Channel
Figure
,
External Mixer
,
, then
Agilent 11970
, and enter a center