Integrating A Signal - Agilent Technologies 1290 Infinity LC System System Manual And Quick Reference

Table of Contents

Advertisement

Integrating a Signal

1 Select the Integration Task Tool (see figure below). The Integrate icon and
2 Click the Set Integration Events Table icon to open the table as shown.
3 Set Baseline Correction to Advanced for gradient runs.
4 Set Slope Sensitivity to 50. Higher numbers will integrate steeper peaks and
5 Set the Peak Width value to the narrowest peak of interest, approximately
6 Area Reject and Height Reject can be set to reject the smallest peaks.
7 Click the Integrate icon to update the results using these new settings.
Agilent 1290 Infinity LC System Manual and Quick Reference
the Set Integration Events Table icon are highlighted in the figure shown below.
ignore less steep peaks.
0.01 in this case.
5
Quick Start Guide
Data Analysis
105

Advertisement