Replacement of detecting unit
(1) Remove the electronics unit according to "Replacement of electronics unit"
(2) Remove the 2 hexagon socket bolts from the electronic housing.
Pull the electronics housing straight forward and away from the detecting unit.
(3) Replace the detecting unit with a new one of the same type.
(4) Check if the O-ring of the cell neck and all the surfaces in contact between the electronics
housing and the cell neck not damaged. Assemble measuring cell and electronics housing.
Tighten the 2 screws
(7) After assembly, perform zero and span adjustment
PROHIBTION
TN5FCXA2V5h-E
Hexagon socket bolt
Amplifer
as the original by comparing nameplates.
• When removing the transmitter case, pay attention not to damage
Detecting unit
Flatcable
95