Fuji Electric FCX-AII-V5 series Instructions Manual And Service Instructions page 58

Hide thumbs Also See for FCX-AII-V5 series:
Table of Contents

Advertisement

M
M
56
%
SPAN
M
%
SPAN
M
%
SPAN
M
A-2 Span (URV) calibration
• Press the
key on the screen
calibration mode.
The measured value and unit on the screen ( ) are
"SPAN" light up.
• Apply the actual input pressure on the screen
ter checking the measured value, press the
• "SPAN" blinks on the screen
the screen
to perform a span calibration at the in-
put pressure at the time.
To perform a span calibration at a point other than
100%, input an appropriate set value (%) ( ) with
the
and
keys, and press the
Setting range
• Select whether the span calibration value setting is
saved on the screen
Press the
key to save the span calibration value
setting and return to the screen
Press the
or
key to cancel the setting and return
to the screen
.
• Check that the span calibration was performed as in-
tended.
Press the
key to perform a span calibration again.
Press the
or
key to move to the next screen for
item name selection.
to select the span
key.
. Press the
key on
key.
.
.