7.3 Replacement of defective parts

the transmitter and bring it into the instrumentation workshop.
DANGER
failure is detected, please replace the concerned part.l
(1) Turn off the power supply.
(2) Remove the indicator.
(4) Unplug each connector carefully.
housing and connect the connectors.
Assemble all parts in the reversing way than de-
scribed before.
Check all the screws are correctly tightened, and screw the electronics housing cover on
the transmitter
PROHIBTION
PROHIBTION
(6) After completion of replacement, perform zero and span adjustments.
94
When removing an explosion-proof transmitter, turn OFF the main
power, then disconnect the piping and wiring.
Do not remove it when the power is ON to prevent serious accident
When installing the electronics unit, make sure that
the zero adjust screw end is in front of the groove in
The electronics unit should be removed carefully so as not to
damage the internal wiring.
Indicator
Amplifer
Zero adjustment
Groove
TN5FCXA2V5h-E