M
M
64
%
SPAN
M
%
SPAN
M
%
SPAN
M
I-2 URV Adj : Span adjustment by changing the
range (URV) (URV adjustment)
• Press the
key on the screen
adjustment mode.
The measured value and unit on the screen
"ZERO" light up.
• Apply the the reference pressure for URV (20 mA) on
the screen
. After checking the measured value
corresponding to 20 mA, press the
• "SPAN" blinks on the screen
the screen
to perform a span (100% point) adjust-
ment at the input pressure:
(upper limit)
• Select whether the URV adjustment value setting is
saved on the screen
Press the
key to save the URV adjustment value
setting and return to the screen
Press the
or
key to cancel the setting and return
to the screen
.
• Check that the span adjustment (URV) was per-
formed as intended on the screen
Press the
key to perform a span adjustment again.
• Press the
or
key to move to the next screen for
item name selection.
PROHIBTION
If the input-output is adjusted, the measurement
range is changed as shown automatically in the fol-
lowing page.
LRV adjustment
The measurement range (LRV and URV) are
changed. The span is not changed.
URV adjustment
Only the URV (span) of the measurement range is
changed. The zero point (LRV) is not changed.
The following are the setting conditions for the adjust-
ment point:
per limit)
Note 1: Output adjustment value (%) corresponding
to the input pressure for the LRV adjustment
Note 2: Output adjustment value (%) corresponding
to the input pressure for the URV adjustment
to select the URV
key.
. Press the