Start Up Of The Fxw - Fuji Electric FCX-AII-V5 series Instructions Manual And Service Instructions

Hide thumbs Also See for FCX-AII-V5 series:
Table of Contents

Advertisement

6.3.2 Start up of the HHC
• Put on/off switch of the HHC on "ON" position.
Put on the enclosed "key" in the corresponding location of the HHC. Without the key and
with the key in vertical position, you can just read the transmitter parameters.
To write new parameters in the transmitter, the key needs to be in horizontal position. Othe-
rwise, you will have an inscription on the HHC screen "INHIBIT KEY OK ?" to let you know
that the key needs to be turned to enable the programming of new parameters in the trans-
mitter.
Nota : "INHIBIT KEY" means that the key permits or inhibits (prohibited) writing parameters
in the transmitter
• The transmitter version and the revision of the HHC software are indicated on the screen
during the start up.
After around 4 seconds the inscription "PUSH MENU KEY" appears (please push menu key)
HHC with the optional printer will have the inscription "PAPER FEED". Please push on <INC
key. By pushing on <INC
"PUSH MENU KEY" will be indicated on the screen by if you push on the clear <CL
On the screen appears the inscription "RECEIVING START". The HHC reads out the data
In case of a connection problem, "NO CONNECTION" will appear on the screen. The "PUSH
MENU KEY" appears again if you push the clear <CL
The reasons of a communication problem can be :
- The 4-20 mA output is not powered.
- The 4-20 mA is disconnected.
supply.
- The HHC is not connected to the correct terminals
steps are indicated on the bottom line of the screen inside following signs ( <
N+1), or the DEC (decrease
The <CHNG
the letter. To add a space between caracters, you have to push the keys <ALHA
To delete caracters, please use the clear key <CL .
a second time by replying on the question "CHNG OK ?
time on the enter < ENT
At this moment the new programmed information are written in the transmitter memory,
"WRITE" indication will appear on the screen of the HHC
TN5FCXA2V5h-E
INC (increase
The most important menus can be
ALHA
ENT
71

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents