48
M
M
M
M
Menu 6 : Measuring range (LRV, URV)
LRV: Lower range value (0% point)
URV: Upper range value (100% point)
ZERO
Selectable setting range
ZERO
Note) If the set value of the LRV is outside the range,
a setting error also occurs in the URV setting, and
vice versa.
The maximum setting range is ±99999.
The URV may exceed the upper limit depend-
ing on the change of the UNIT. If that happens,
ZERO
6-1 : Change of LRV (lower limit of the measuring
• Press the
screen for setting the zero point range (0% point)
(
).
• Input the numerical values with the
the screen
Functions of the keys:
key:
key: To increase the value.
Note)
If "setERROR" is displayed as a LRV value, the
range value is not supported.
• To set the decimal point position, press the
the screen
name (
) and you can set the decimal point position
with the
and
key: To move the decimal point position to left
key: To move the decimal point position to right
• Select whether the LRV setting is saved on the
screen
.
Press the
Press the
key on the screen
.
To decrease the value.
. "P" is displayed at the left of the unit
keys.
key to save the zero point range setting.
or
key to cancel the setting.
TN5FCXA2V5h-E
to display the
and
keys on
key on