Fuji Electric FCX-AII-V5 series Instructions Manual And Service Instructions page 91

Hide thumbs Also See for FCX-AII-V5 series:
Table of Contents

Advertisement

INC
1
J:
SATURATE CUR
< 1 >SAT LO Y.YmA
< 2 >SAT HI YY.YmA
< 3 >SPEC XXXXXX
1
INC
1
SATURATE LO
2 SATURATE HI
3 SPEC
SATURATE LO
(Lower limit value of
saturation current)
Y.YmA
4.0mA
3.9mA
3.8mA
3.7mA
INC
DEC
3.6mA
3.5mA
3.4mA
3.3mA
3.2mA
Press <INC> or <DEC> to change
the saturation current value.
(See above)
INC
3
J:
SATURATE CUR
< 1 >SAT LO Y.YmA
< 2 >SAT HI YY.YmA
< 3 >SPEC XXXXXX
1
INC
SPEC
Menu No.K
NORMAL
INC
DEC
EXP.
XXXXXX part:
NORMAL=Standard spec.
EXP.=Extended spec.
TN5FCXA2V5h-E
J-1: SATURATE CUR
SATURATE LO
Y.YmA
< CHANGE > < CL >
CL
CHNG
CL
J-2: SATURAT CHNG
SATURATE LO
Y.YmA
< INC > < ENT> < CL >
CL
ENT
J-2: SATURAT CHNG
SATURATE LO
Y.YmA
CHNG OK?< ENT/CL >
J-1: SATURATE CUR
SELECT SPEC
XXXXXX
< INC> < ENT> < CL>
CL
ENT
CL
J-1: SATURATE CUR
SELECT SPEC
XXXXXX
CHNG OK?< ENT/CL>
Menu J : Saturation current value and
2
Change of saturation current value (Lower limit)
(Changeable only on "SPEC EXP")
3
cancel
4
but above the burnout current (UNDER SCALE)
Change of saturation current value (Upper limit
value)
Make a setting as same as the setting of the lower
limit value by input 2 on display
SCALE)
* Burnout current is settable according to "9. Burnout
direction and value".
5
lectable.
Refer to "J-3" in "J. Saturation current value and
play for details.
6
-
.
-
-
89