Menu A : Zero/span calibration
This menu gives the poissibility to calibrate Zero (LRV)
and span (URV) of the transmitter.
The ranging of the transmitter needs to be done in the
"Range" menu 6. PLease use menu A "CALIBRATE"
menu.
It is absolutely necessary to the zero or LRV point as
welle as the span or URV point when making a cali-
bration.
INDICATION
1. After performing a zero calibration, perform a
span calibration.
2. If you input the value that exceeds the adjustable
range, the setting will not be changed even after
the setting is saved.
Adjustable range
Zero calibration: within ±40% of the max span
Span calibration: within ±20% of the set span
A-1 Zero (LRV) calibration
• Press the
key on the screen
calibration mode.
The measured value and unit on the screen ( ) are
"ZERO" light up.
• Apply the actual input pressure on the screen
ter checking the measured value, press the
• "ZERO" blinks on the screen
the screen
to perform a zero calibration at the in-
put pressure at the time.
To perform a zero calibration at a point other than 0%,
input an appropriate set value (%) ( ) with the
keys, and press the
• Select whether the zero calibration value setting is
saved on the screen
Press the
key to save the zero calibration value
setting and return to the screen
Press the
or
key to cancel the setting and return
to the screen
.
• Check that the zero calibration was performed as in-
tended.
Press the
key to perform a zero calibration again.
Press the
or
key to move to the next screen for
item name selection.
to select the zero
key.
. Press the
key on
key.
.
.
. Af-
and
55