ELECTROMAGNETIC COMPATIBILITY

ELECTROMAGNETIC COMPATIBILITY
All FCX-AII series of pressure transmitters are in conformity with the provision of the EMC Directive 2014/30/EU on the
harmonization of the laws of the Members States relating to electromagnetic compatibility.
All these models of pressure transmitters are in accordance with the following harmonized standards :
• EN 61326-1 (Electrical equipment for measurement, control and laboratory use - EMC requirements -
Part 1: General requirements).)
• EN 61326-2-3
Emission limits (according to EN 55011 / CISPR 11, Group 1 Class A)
Frequency range (MHz)
30 to 230
230 to 1000
Immunity
Phenomenon
Electrostatic Discharge
Radiated, Electromagnetic
Field
Fast transients (burst)
Surge Transients
Conducted RF Disturbances 3 Vrms (150 kHz to 80 MHz) EN/IEC 61000-4-6
Power Frequency
Magnetic Field
Performance criteria (A & B): according to IEC 61326
TN5FCXA2V5h-E
Limits
40 dB (μV/m) quasi peack, measured at 10 m distance
47 dB (μV/m) quasi peack, measured at 10 m distance
Test value
±4 kV (Contact)
±8 kV (Air)
10 V/m (0.08 to 1.0 GHz)
3 V/m (1.4 to 2.0 GHz)
1 V/m (2.0 to 2.7 GHz)
2 kV (5/50 ns, 5 kHz
1 kV Line to line
2 kV Line to ground
80% AM @ 1 kHz
30 A/m (50 Hz, 60 Hz)
Standard
Performance criteria
EN/IEC 61000-4-2
EN/IEC 61000-4-3
EN/IEC 61000-4-4
EN/IEC 61000-4-5
EN/IEC 61000-4-8
Result
Passed
Required
Result
of criteria
B
A
A
A
B
A
B
A
A
A
A
A
3