6 Automatic Access Test
If both modes (P-P / P-MP) are found when Layer
2 on the D-channel is checked, the mode must be
selected manually (see Page 31).
If everything has been correctly detected, the ARGUS will
display the successfully found access and access mode in
the third line. Additionally, a qualitative assessment of the
level will be displayed.
The ARGUS will automatically determine the protocol or use
the manually selected protocol (see Page 116) . On a
bilingual access, the ARGUS will use the DSS1 protocol.
LED L3 will light after the ARGUS has setup Layer 3.
At the same time the ARGUS will start the B-channel test.
The results will be displayed on the ARGUS. If an error
occurs in the B-channel test (e.g. access is not plugged-in),
the ARGUS will - depending on the class of error - either
repeat the initialization or show an error message (see
ARGUS Error Messages Page 144).
The ARGUS will then idle in the Status display.
Status display example:
Softkeys:
The current settings of the softkeys are displayed in the
fourth line.
The ARGUS is in largest part operated with the two ↓ ↓ ↓ ↓ ↑ ↑ ↑ ↑ -
Keys, the confirmation key  and the three softkeys.
On the following pages, only the softkey's meaning in the
respective context is shown - enclosed in brackets < > ,
e.g. < NO >.
The <  > softkey serves the same function as the 
22
The ARGUS displays the
type of access (BRI), the
availability of the B-channels
(B12),
a level evaluation
(OK),
the access mode (TEs), the
bus
configuration
and the protocol (DSS1).
If you press the <RESTART>
softkey, the B-channel test
will be repeated.
ARGUS 3u plus/3u NT
(P-MP)