Chapter 5 Installation Procedures; Insert The Perforated Shelves; Preparing The Useable Space; Mains Connection - Thermo Scientific Heratherm IMC 18 Operating Instructions Manual

Table of Contents

Advertisement

Installation procedures
Remove the packing foil.

Insert the perforated shelves

Remove the transport securing devices from the perforated shelves.
Insert the perforated shelves in the grooved profiles on the side walls of the useable space.

Preparing the useable space

The Heratherm Incubator IMC 18 is not delivered in a sterile condition.
The following useable space components should be checked for cleanliness and disinfected prior to use:
• perforated shelves,
• useable space surfaces,
• door seal,
• useable space door.

Mains connection

Connect the device to the on site fused mains supply with the following values:
• Safety fuse (time-lag) T 16 A or circuit breaker B 16.
• Connection via a fault current circuit breaker (triggering current < 30 mA) is recommended.
Thermo Scientific
Contact with live electrical components may cause a lethal electric
shock.
Before connecting the device to the power supply, check the power
cord and the plug for damage. Do not use damaged cables for con-
necting the device to the power supply!
Electrical shock
Heratherm IMC 18
5
5-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents