3.3.3 Environmental

Temperature, operating: 21°C ± 5°C (70°F± 9°F)
Temperature, storage: 5°C
Humidity, operating: 10
Mounting surface: Sturdy, stable, capable of handling 128 kg (283 lbs)
weight
Pollution degree: 2
Installation category: II

3.3.4 Electrical

There are two electrical configurations of the Echo liquid handler:
Config. Voltage
A
100–120 VAC
B
200–240 VAC

3.3.5 Client PC

Operating system: Microsoft
using 32-or 64-bit OS; Windows Net Framework 2.0
®
CPU: Intel
Core™ 2 Duo or later
Memory: 1GB or greater
Hard drive storage: 1GB
Video: 1280 x 1024 resolution
API: ActiveX
Programming languages: C, C++, C#, Visual Basic, Visual J#, or
Windows Net Framework compatible

3.3.6 Communications

Network connection: 10/100/1000 BaseT
Network protocol: TCP/IP
User account: Labcyte (The client PC must have a Labcyte account, with
administrator privileges, that is accessible by LabcyteService and Support to
service the Echo liquid handler system.)
45°C (41°F
113°F)
80%, non-condensing
Frequency
Power Fuse
50/60 Hz
10 A
50/60 Hz
5 A
®
Windows XP SP3 or Windows 7
250 VAC, 10 A, Fast Blow
250 VAC, 5 A, Fast Blow