C.3
Chiller Setup and Operation
C.3.1 Setup
Your Labcyte service representative will set up the chiller during Echo
system installation. If you need to set up the chiller again (for example,
after relocation of the Echo system), use the following procedure.
The Echo system should already be set up, with tubing assembly attached
to the fluid panel (see Echo liquid handler user manual).
1. Place the chiller on the benchtop near the Echo unit, or on the floor
2. Connect the Echo tubing assembly to the fluid connections on the chiller
Figure C.2 Tubing assembly connections
3. Connect the power supply cable:
under the Echo unit.
labeled Coolant Return for liquid returning to the chiller and Coolant
Supply for fluid exiting the chiller.
circular 2-pin connector to the 12 V DC (8 A) on the chiller.
three-prong electrical connector to the electrical outlet.
C-3