Labcyte Echo 520 User Manual page 125

Liquid handler
Table of Contents

Advertisement

Barcode Location : From the drop down list, select the barcode location.
The choices are N
ONE
supported.
Note: Some users are more familiar with the microplate sides
identified by the compass points: L
south.
Figure 5.27 Selecting barcode locations
Left or West
Left or west
position
position
Plate dimensions
Two illustrations identify the destination plate dimensions that must be
entered. These dimensions may already be specified in the SBS standards
(see
"SBS standards" on page
Figure 5.28 Plate dimensions A, B, C, and D
A and B: Well A1 X/Y offsets. Dimensions A and B locate the center of
well A1 in relation to the plate's left and top outside edges. A specifies the
A1 well center to the left outside edge in millimeters. B specifies the A1
well center to the top outside edge in millimeters.
Examples: For a 1536LDV DMSO source plate type, A is 11.015 mm and
B is 7.94 mm.
The valid range for A is 0.00 mm to 128.00 mm.
The valid range for B is 0.00 mm to 86.00 mm.Refer to ANSI/SBS 4-2004
for more information.
, L
-S
and L
EFT
IDE
ONG
EFT
Long, front or south
5-27).
-S
(front). The right side is not
IDE
-S
= west and L
IDE
ONG
Echo Liquid Handler Software 5-25