Power
2. Click the C
P
procedure:
5-48 Echo Liquid Handler User Manual
Amplifier Feedback: The A
current power correction factor, and normally should be very close to
1.0.
Pulse Energy: The P
calibration is performed.
Vpp: The voltage peak-to-peak (V
calibration is performed.
button in the P
ALIBRATE
dialog box. The Echo liquid handler performs the following
OWER
The most recent power calibration record stored in the Echo database
is read. This data will be used in the next step.
An RF pulse is generated.
Several RF measurements are taken.
A new amplifier feedback value is calculated and displayed as
P
—A
OWER
MPLIFIER
A new pulse energy value is measured and displayed as P
P
E
.
ULSE
NERGY
A new Vpp value is measured and displayed as P
The new power calibration record is stored in the Echo database.
F
MPLIFIER
EEDBACK
E
field will be empty until power
ULSE
NERGY
) field will be empty until power
PP
section of the C
OWER
F
.
EEDBACK
field displays the
ALIBRATE
OWER
—V
.
OWER
PP