If register calibration is required, use the following procedure:
1. Click R
window opens.
Figure 5.42 Plate registration window
2. Select the source plate type from the drop-down menu.
3. Click the P
4. Place the source plate on the plate gripper.
5. Click the R
6. Click the R
position and displays the new well center values.
7. Click the A
Y
Figure 5.43 Plate registration confirmed
5-46 Echo Liquid Handler User Manual
in the Calibration window. The Plate Registration
EGISTER
button to extend the source plate gripper stage.
RESENT
button to draw the plate into the instrument.
ETRACT
button. The Echo liquid handler calibrates the well
EGISTER
L
PPLY
ABWARE
in the Confirmation Update dialog box, to update the database.
ES
button at the bottom of the dialog box, then