6.1

Maintenance schedule

Echo user maintenance consists of the following:

6.1.1 Daily maintenance

Perform the following checks once a day:
Maintenance Task
• If the coupling fluid level is at 350 mL or less,
• If the coupling fluid is cloudy, clean and refill
• Check the waste bottle daily (older Echo
6-2 Echo Liquid Handler User Manual
Daily maintenance
Echo maintenance alerts
Scheduled preventive maintenance
Every two weeks
Every month
Every six months
Table 6.1. Daily Maintenance Checklist
Check the coupling fluid bottle level daily.
clean and refill the coupling fluid bottle and
add algaecide.
the coupling fluid bottle and add algaecide.
systems only). Empty it if it is greater than
half full.
Check the plate insert for proper positioning
or damage.
Procedure
See "Refill the coupling fluid
bottle" on page
6-8.
See "Clean and refill the
coupling fluid bottle" on
page
6-10.
See "Empty and clean the
waste water bottle" on
page
6-13.
See "Source plate insert" on
page 3-6