5. Transfer Volume: Enter the total volume of fluid to be transferred. The
valid range is dependent on the source microplate (see
Transferred Volume:" on page
384-well PP source microplate, the valid range is 2.5 to 10000 nL in
increments of 2.5 nL
using the increment/decrement buttons, remember that the value must
be an even multiple of the minimum transfer volume. If you type an
invalid transfer volume, you will receive an error message that displays
the correct transfer range.
Figure 5.6 Invalid transfer volume for 384-well plate
6. Click the F
button to complete the new protocol.
INISH
Figure 5.7 384:384 protocol finished
* The valid range may also differ for different fluid types.
3-21). For example, for the Labcyte
*
. If you enter the transfer volume directly instead of
"Maximum
Echo Liquid Handler Software 5-7