Labcyte Echo 520 User Manual page 112

Liquid handler
Table of Contents

Advertisement

Figure 5.14 Add new protocol wizard page 1 of 3
3. Click the N
box opens.
Figure 5.15 Add new protocol wizard page 2 of 3
a. Transfer Map Type: Select an interleaved or a quadrant liquid transfer
b. Number of Copies: Enter the number of times the protocol will be
Transfer Volume: Enter the transfer volume in 2.5 nL increments up to
2000 nL or up to 500 nL for aqueous transfers. If you enter the transfer
volume directly instead of using the increment/decrement buttons,
remember that the value must be an even multiple of 2.5 nL.
4. Click the N
opens.
Figure 5.16 Add new protocol wizard page 3 of 3
5-12 Echo Liquid Handler User Manual
button. The second A
EXT
protocol type.
See "What types of liquid transfer are supported?" on
page
1-12.
executed. The valid range is 1 to 10,000 copies. You are prompted to
load and unload the source microplate and the specified number of
destination microplates.
button. The third A
EXT
N
P
DD
EW
ROTOCOL
N
P
W
DD
EW
ROTOCOL
W
dialog
IZARD
dialog box
IZARD

Advertisement

Table of Contents