Labcyte Echo 520 User Manual page 177

Liquid handler
Table of Contents

Advertisement

6. Turn on the Echo carriage motors by turning the Emergency Off switch
counterclockwise and releasing it.
Clean the destination plate gripper grid
The grid on the destination plate gripper is charged with an electrical
current during fluid transfer. The electrical field overrides any static charge
that may reside in the destination plate to ensure that the droplets
transferred from the source plate wells are directed upward into the center
of the destination wells and not diverted to the side of the wells.
The grid should be inspected for particles and cleaned monthly.
Tool needed
Compressed air or vacuum with brush
To clean the destination plate gripper grid:
1. Display the Diagnostics window in the Echo software.
2. Click Door Open to open the process door.
3. Click Destination Plate Out. The software extends the destination
plate gripper and automatically turns off the electrical current.
Figure 6.8 Destination plate gripper extended
4. Gently remove any dust or particles on the grid with compressed air or
vacuum with brush.
5. Click Flip Destination Plate Down and repeat the previous step to
clean the grid.
6. Click Destination Plate In and select None from the Plate Name drop-
down menu in the Retract Dest Plate Gripper dialog box. The software
retracts the destination plate gripper.
Maintenance and Service 6-17