5.3.1 Device control

You have access to precise Echo device control. The ability to command
internal mechanism movement is useful during maintenance,
troubleshooting, and diagnostic procedures. Access to device control is
found in several areas of the Echo software:
The Device Control and Status section of the D
uses eleven sets of checkboxes and status indicators. They are
discussed below.
The Miscellaneous section of the D
diagnostic procedures drop-down list and L
See "Diagnostic procedures" on page
The M
section of the C
OTION
R
buttons.
EGISTER
Note: The Echo liquid handler will not allow device movement that
would cause damage to itself. For example, if the process door is
closed when the source plate gripper stage is commanded to extend
outside, the Echo liquid handler opens the process door first.
Figure 5.32 Device control and status section
Door open/close
The Echo door can be manually opened and closed by selecting the
appropriate checkbox.
If either the source or destination plate
gripper stage is extended outside of the
Echo liquid handler when the process
door is commanded to close, a "Close
Door Error" error message is displayed.
The associated indicator lights report the current position of the door.
IAGNOSTICS
ALIBRATION
See "Motion calibration" on page
IAGNOSTICS
window uses the
button.
AUNCH
5-38.
window uses the H
5-45.
Echo Liquid Handler Software 5-29
window
and
OME

Advertisement

Table of Contents
loading

This manual is also suitable for:

Echo 550Echo 555

Table of Contents