5.3.2 Status indicators

The S
S
YSTEM
TATUS
array of status indicators and readouts. Four of them (P
F
, and EMO) correspond to the similarly named indicators on the Echo
AULT
status panels.
Figure 5.33 System status section
Power indicator
The P
status indicator provides more information than simply power
OWER
on or power off.
Green: The Echo liquid handler is powered on and is ready for use.
Gray: One or more of the following conditions is true:
The Echo liquid handler is not powered on.
The Echo internal controller is not running.
The Echo liquid handler is still initializing after being powered on and
is not yet ready for use.
Warning indicator
The W
status indicator reports conditions that will not allow normal
ARNING
operation.
Gray: OK. The Echo liquid handler is ready for use.
Yellow: One or more of the following is true:
The Echo stages are not homed.
page
5-45.
The coupling fluid bottle level is low.
on page
5-35.The process door is neither fully open nor fully closed.
The upper anti-static bar is neither fully up nor fully down.
The source plate dryer nozzle is neither fully up nor fully down.
Focus calibration is required.
section of the D
IAGNOSTICS
See "Home calibration" on
window contains a vertical
, W
OWER
See "Coupling fluid indicator"
Echo Liquid Handler Software 5-33
,
ARNING